Possible Causes: If liquid has been manually poured into the vessel before the computer communicates with the appropriate valve to fill the vessel, this fault screen will appear. Open the vessel, if the vessel is full press "Y<ENTER>" to continue the...
The TDF instrument produces about 475 ml per assay of filtrate waste consisting of about 75% ethanol-based solution, about 2,850 ml per run (6 stations). There is also water, buffer solution, and ethanol pumped through the system to flush lines and...
Over a period of years, the buffer valve, and possibly others, may begin to corrode the brass valve body. As of November 2022, the buffer valve in new instruments comes with a PEEK chemical-resistant valve body. However, older instruments may show...
1. Verify that you have a Gen2 Coordinator: A Gen2 Coordinator will have a number that is less than 1,000. If you have connectivity issues and your coordinator number is greater than 1,000 you have a Gen3 system. In that case, contact ANKOM using the...
The F57 and F58 Filter Bags are ash-free and therefore can be used for ash determination. Refer to the Ashing Procedure Service Procedure (AS034) linked below.
The Dairy One Forage Laboratory receives samples of feed, forages, byproducts, and ingredients from around the world each day. Known for their versatility of livestock and exotic animal feed analysis across a variety of industries, Dairy One needed...
Located at the picturesque Cochno Farm and Research Centre as part of the University of Glasgow, the lab team of Prof. Murray, Prof. Hastie, Dr. Farci, Daisy Taylor, and Stefan Yerby specialize in equine and ruminant nutrition. Together, they...
Refer to Service Procedure Insert Samples Into F58 Bags (AS032).
Possible Causes: The Q6 Valve (drain valve) is not opening correctly or there is a line blockage between the Q6 valve and the vessel. To best assist you, run a ~10 minute test. Please contact ANKOM with your results. To test: Verify that the Q6...
To minimize bag material abrasion, use the new bag suspender (as of March 2015 - Part #: F11 ). It has been redesigned to include rounded bottom points on each tray.
The XT instrument’s flowmeter should have a flow rate between 4-6 gallons per hour (If chilled coolant is used then the flow rate should be 2-3 gallons per hour). Note that this flow is intermittent and only turns on when cooling is required. When...
Yes, all samples should be treated the same during the drying process. The biggest and most common cause of artificially high crude fat results is incomplete drying in the pre-extraction phase. Because our technique uses the indirect method of...
The valve will stay open for 90 seconds. The valve(s) can be closed before this time by clicking the Valve Open check box (to remove the check mark from the box), clicking the Close Valves button, or clicking the R ecord button. Since opening the...
When you press the record button on your GPM software screen, the program automatically stores your data in the file shown in the AutoSave location box. Therefore, if there has been a computer interruption during a study, you will be able to retrieve...
Refer to the Instrument Software Functionality section of the RF Gas Production System Operator's Manual.
Possible causes include: A loose or disconnected wire from the pressure sensor A faulty circuit board To diagnose, unplug the pressure sensor connector from the circuit board and plug in a pressure sensor connector from a working module into the same...
The vessel temperature has exceeded the maximum limit. Possible Causes: The solution level is low in the vessel, but the level sensor is indicating full. The level sensor has two main components that work together. These components are a Level Sensor...
The instrument is experiencing a high-temperature fault. If your controller is set to 100°C, there are two possible reasons for this fault. (1) The agitation is not working properly during an assay. To check the agitation during an assay, remove the...