An E11 fault is the result of the high-pressure sensor reading input pressure below 35 psi. N2 input pressure is always monitored and the fault could occur at any point during a run. The fault will interrupt the run and wait for the operator to...
The Silicone Pump Tubing ( Part #: TDF68 ) should be changed when the instrument prompts the user with an E13 warning for weak pump tubes. The pump tubes typically last 6 months under normal use. Heavy usage may necessitate a more abbreviated...
Yes. Please refer to TDF Service Procedure Return Shipping Instructions (TS019) for details.
Static electricity during the weighting process: The largest contributor to poor results is the effect of static electricity on the weighing process. Because plastic gloves can contribute to static electricity, only use anti-static gloves. To...
To properly prepare the IDF & SDF bags for Kjeldahl and Ashing procedures, refer to Section 17 of the TDF Operator’s Manual (Ash Determination) or view the following video: Kjeldahl & Ashing Bag Preparation .