ANKOM HCl Hydrolysis System Operator's Manual
When the rotation switch is activated on the Daisy, the jars should continue to rotate until the switch is turned off. Possible Causes: There is not enough friction between the outside surface of the jars and the rollers on the instrument. This is...
Contact ANKOM Technology via e-mail at https://www.ankom.com/contact/technical-services or by calling 1-315-986-8090 for assistance in making this determination. Follow D200 Service Procedure 1 to replace the motor on the Daisy instrument.
This may be the result of one of the IDF or SDF valves being stuck in a closed or partially closed position. Using the Diagnostics screen select Valve Test and look for IDF Inlet, SDF Inlet , and Waste buttons. Run a brief test using the screen to...
Refer to TDF Service Procedure Long-Life Pump Tubing Replacement (TS028) for instructions.
Yes, there is a recommended procedure for the TDF Instrument for periods of non-use in Appendix C of the Operator's Manual for instructions. Following this procedure will help maintain clear fluid lines. The most current Operators Manual can be found...
Possible Causes: The rinse water supply temperature did not meet method standards. This may affect results with certain samples. Make sure the water supply has a temperature of no less than 50°C for Crude Fiber or no less than 70°C for ADF and NDF...