Possible Causes: The controller needs to be reset. See DELTA Service Procedure Controller Temp Menu Setup (DES015) to determine if menu settings are incorrect and to make the appropriate changes. If the controller reset doesn't resolve the problem...
Possible Causes: Faulty DELTA Temp Controller ( Part #: A11 ). Refer to DELTA Service Procedure Controller Temp Test (DES014). Faulty Splat Controller ( Part #: A09 ). Poor connection between the DELTA Temp Controller and the Splat Controller.
Possible Causes: (1) The controller needs to be reset. Refer to Service Procedure Controller Temp Menu Setup (AS015) to determine if menu settings are incorrect and, if so, make the appropriate changes. (2) The RTD may need to be replaced. Contact...
Possible cause: A power surge has occurred. Solution: Reprogram your controller using the controller menu setup in your Operator's Manual
The instrument is experiencing a high-temperature fault. If your controller is set to 100°C, there are two possible reasons for this fault. (1) The agitation is not working properly during an assay. To check the agitation during an assay, remove the...
Possible Causes: The solution temperature has exceeded the maximum limit. Perform the boil test detailed in Service Procedure Boil Test (AS013).
The instrument is experiencing a pressure fault. There are three possible reasons for this fault. (1) More than 2 L of solution was in the vessel during the digestion process. A pressure fault will most commonly occur if there is more than 2 L of...
"----" indicates there has been an electrical malfunction. Reset the Controller. Refer to A2000 Service Procedure Controller (Temp) Menu Setup (AS015). If the "----" still shows, you will need to replace the Ascon Controller ( Part #: A09 ). Possible...
Possible Causes: Faulty A2000 Temp Controller ( Part #: A11 ). Refer to Service Procedure Controller Temp Test (AS014). Faulty Splat Controller ( Part #: A09 ) Poor Connection between the Temp Controller and the Splat Controller (AS020).
"----" indicates there has been an electrical malfunction. You will need to reset the Controller. Refer to DELTA Service Procedure Controller Menu Setup (DES015). If the "----" still shows, you will need to replace the Ascon Controller ( Part #: A09...
To ensure that the controller settings are correct, refer to A200 Service Procedure Controller Reset Procedure (FS002) for instruments purchased pre-9/03, or, see A200 Service Procedure Controller Menu Setup (FS001) for instruments purchased post...