|
Troubleshooting Guide
Your ANKOM instrument has been designed for long service and ease of
maintenance. Most problems are easily diagnosed and fixed. If you experience
problems that are not addressed in the information below or if you have
any questions, contact ANKOM Technology for assistance. We are committed
to fast and reliable service.
Problem: The Fiber Analyzer will not turn on.
Possible Causes:
The power cord is not plugged into the unit or the power source.
The main power switch is not turned on.
One of the system circuit breakers needs to be reset.
Problem: Fluid is observed in the Reservoir of the
Maintenance Alert.
Possible Causes:
Leakage has occurred at the agitator shaft.
The Packing Nut inside the system requires attention.
Disconnect the power supply and remove the back of the instrument. Visually
inspect the agitator assembly, specifically the area between the Maintenance
Alert Collector and the Packing Nut. If signs of leakage are evident,
clean the area thoroughly and tighten the Packing Nut 1/8 to 1/4 turn.
If overtightened, the motor will sound strained or will not turn. If this
occurs, loosen the Packing Nut slightly until the motor turns freely.
Check the agitator assembly after the next procedure to ensure a tight
seal. If no leakage is evident, thereafter, check the reservoir periodically.
If leakage is evident repeat Step Two. Every 20 hours of operation check
the agitator assembly for signs of leakage. When the instrument will not
be in use for a long period of time, fill the vessel with water and run
the agitator for approximately 10 minutes. Rinse the vessel thoroughly
out with water. This will ensure no acid residue will corrode the instrument
during storage.
Problem: The agitator is not moving and/or the motor
sounds strained.
Possible Causes:
The Packing Nut may have been tightened too much during periodic maintenance
or during operation.
Try loosening the Packing Nut slightly until the motor begins to operate
and agitator moves freely.
If periodic maintenance was not accomplished and acid solution was allowed
to leak into the packing assembly the shaft may become "frozen."
Check for corrosive buildup. The Packing Set and Agitator may need to
be replaced.
Problem: After some time, the temperature settings
and the actual temperature do not match. The actual temperature is higher
or lower than the factory setting.
Possible Causes:
The temperature controller is malfunctioning.
The temperature probe is malfunctioning.
Carefully pull the temperature controller out of the instrument by grasping
both sides and pulling straight out. Wiggling from side to side may be
necessary. Firmly re-insert the controller. Pour about 2000 ml of water
into the vessel. Leaving the vessel lid open, turn on the heat and agitate
switches and allow the system to come to a boil. NOTE: The water
should begin to boil in about 10 minutes. Do not let the water boil over.
Once the water begins to boil, check the temperature reading. The temperature
readout should be at approximately 100 ° C. If the readings do not
match, contact ANKOM Technology. Record the results of your boil test
to include, time to boil and temperature reading at boil.
Problem: The timer no longer has a readout.
Possible Cause:
The timer battery needs replacement.
See the Timer Manual that was shipped with your instrument.
Problem: Temperature readings on controller fluctuate.
Possible Cause:
Contamination on controller contacts.
Carefully pull the temperature controller out of the instrument by grasping
both sides and pulling straight out. Wiggling from side to side may be
necessary. Clean contacts at the back of the controller. Firmly re-insert
the controller.
For problems relating to the results of your analyses using the
ANKOM Fiber Analyzer first check to ensure the procedures sent with your
machine were followed. There are also copies of the major procedures
on-line. Some possible variables that may affect your results are:
Improper solution concentrates.
Some chemicals may have come out of solution
in your mixes and need to be heated and re-mixed.
Rinse water was not heated to 90-100º Celsius.
Oven drying temperatures were too high (drying
temperatures higher than 105º Celsius may affect results).
Improper calculation of blank bag correction or
using the wrong values in the formula.
Modifying the settings on the temperature controller.
If problems persist, please click here to fill out a trouble
report.
|